Produkter » Radioaktivitetsmätning » Handinstrument » Nuklididentifiering » Identifinder R300

Identifinder R300

Request a quotation
  • Small instrument based on CZT-detector, resolution < 3,5%
  • Transflective color display, Easy to read display, even in bright sunlight
  • 24 hour battery life
  • IP67
  • ANSI N42.48 compliant
  • Ease of use in high background environments
  • Highly reliable, accurate
  • Rugged, drop tested 1,5 m

Identifinder R300 features a high performing small sized spectroscopic identification instrument with CZT detector. Like other FLIR radiation identifinder instruments, it has bluetooth, web server and GPS offering full reach back functionality. CZT detectors is more rugged than scintillation-based detectors and does not require temperature stabilization. The R300 is also available with an optional neutron detector as well as a full range of accessories.

Detectors: CdZnTe, Three crystals; 15 × 13 × 5 mm
Neutron He-3; 15mm × 54 mm @ 8 atm
Energy Range: 30 keV – 3 MeV
Energy Range (Gamma); Identification Channel  30 keV – 3 MeV 

Throughput: Dose Rate Channel   ≤250 kcps 
Throughput: Identification Channel  ≤180 kcps
Sensitivity (Cs-137); Identification Channel ≤ 47 cps/µSv/h
Sensitivity (Cs-137): Dose Rate Channel  ≤ 81 cps/µSv/h 
Gamma Spectrum 1024 channels; 3 MeV
Dose Rate Range: 0,1 µSv/h – 10 mSv/h; ±30 % 
Overload Threshold 5 mSv/h
Neutron Sensitivity 2,6 cps/nv; ±20 %
Nuclide Identification: According to ANSI N42.48
Typical Resolution ≤3.5 % FWHM at 662 keV at 25 °C ambient temperature
Recommended Service Interval: 5 years
Dimensions (W × D × H): 70.3 mm  × 33.8 mm × 125.5 mm 
Weight: 340 g without neutron detector  or 370 g with neutron detector
Housing Material: Aluminium
Operating Temperature: -20 °C – +50 °C 
Storage  Temperature: +10 °C – +35 °C
Temperature Change; Temperature change  must not exceed 8.4 °C/min
Relative Humidity: 10 % – 93 %, non condensing
Shock Resistance: According to ANSI N42.48-2008
Protection Rating: IP63 according to IEC 60529
Battery Internal (Li-ion, single cell, rechargeable) Capacity: 1950 mAh; 3.7 V; 25 °C
Battery Life: 300 – 500 full charge-discharge cycles; 3 a – 5 a
Charging Duration :  ≤3 h at 25 °C when instrument off and connected to DC power charger  and >5.5 h when  powered via USB.
Operating Duration:  ≥24 h at 25 °C in dose rate mode with dimmed display back light and GPS switched off
Display Type: Transflective color LCD, 320 pixel × 240 pixel
GPS: integrated
Identifications inclusive Spectrum ≈ 600000
Alarms ≈ 1000000
Input/Output: USB 2.0; mini-B socket or Bluetooth: ≤ 50 m range

Software
Embedded Software: Microsoft Windows CE Operating System
Web Interface: TCP/IP emulated via USB
Download File Formats: ANSI N42.42 and spc files compatible with third-party analysis software
Reachback: ANSI N42.42 data via Bluetooth

Standards
ANSI N42.48: Performance Requirements for Spectroscopic Personal  Radiation Detectors (SPRDs) 
IEC 60529: Degrees of Protection Provided by Enclosures (IP Code) EMC - Directive 2004/108/EC. Regulations  concerning electromagnetic compatibility FCC Part 15. Regulations concerning electromagnetic compatibility
ANSI N42.42 2006: Data format standard for radiation detectors used for Homeland Security Detectors

Accessories
DC Power Adapter, AC in: 100 V – 230 V; 50 Hz – 60 Hz; DC out: 5 V; 2 A 
Car Power Adapter, DC in: 12 V – 24 V; DC out: 5 V; 2 A
Connection Cable Mini USB-B – USB-A; 1.83 m
Holster 60 g
Carrying Case: Included